MACROTEST

5035

EN - 54

If the instrument measures

an

interfering voltage higher than 30V
on the amperometric circuit
, it
doesn’t perform the test and a screen
similar to the one beside appears.

o

31

V

VOL AG


THE PREVIOUS RESULT CANNOT BE SAVED.


If the resistance measured is higher

than the full scale of the instrument,
when pressing

START/STOP the

instrument performs the test and a
screen similar to the one beside
appears.

o

>1999

- - - - - -



If the resistivity measurement is higher

than the following value
1999x6.28x(distance between the rods
selected), when pressing
START/STOP the instrument performs
the test and a screen similar to the one
beside appears.

o> 1999

- - - - - -



SAVING:

The tests can be stored pressing the

SAVE key twice

(according to paragraph 5.1).



Voltage value on the amperometric circuit

SAVE

1999

Ω is the full

scale of the
instrument.

1999

Ω is the full

scale of the
instrument.

SAVE