HT instruments MACROTEST 5035 User Manual
Page 18

MACROTEST
5035
EN - 17
4.2.1.2. Mode
"TIMER"
6. Use the following keys to set the duration time of the test:
Press this key to increase the duration time of the test (
Tmax=999 seconds).
Press this key to decrease the duration time of the test (
Tmin=10 seconds).
7. Press
the
START/STOP key.
The instrument effects the measurement ending when the set time has elapsed.
999 seconds
→ Maximum value of the test duration.
10
seconds
→ Minimum value of the test duration.
Note: Pressing the START/STOP key again the test gets immediately interrupted.
At the end of the test, in case
the
resistance value R
ISO
detected results
to be lower than R
MAX
(depending on
the selected voltage see following table)
and the test is effected at the selected
rated voltage value, the instrument
emits a
double sound signal indicating
the positive outcome of the test
displays one screen similar to the screen
alongside.
R
ISO
2.07
M
Ω
500 V 20
s
TIMER
The values of measured resistance of isolation always must be confronted with the
normative limits for being able to assert if the system is to norm.
SAVING:
The test can be stored pressing the
SAVE key twice
(according to paragraph 5.1).
Note: The maximum resistance value R
MAX
which can be measured in mode M
Ω
depends on the rated voltage selected for the test. In particular:
Rated voltage selected
for the test
R
MAX
= Maximum resistance
value
50VDC
99.9M
Ω
100VDC
199.9M
Ω
250VDC
499M
Ω
500VDC
999M
Ω
1000VDC
1999M
Ω
Maximum resistance values measurable under M
Ω mode depending on the rated voltage selected.
Timer mode.
Duration of the
set test.
Insulation
resistance value
R
ISO.
Rated voltage value selected for the Timer test.
SAVE
START
STOP
S▲
U
L
▼