Caution – HT instruments MACROTEST 5035 User Manual

Page 33

Advertising
background image

MACROTEST

5035

EN - 32

4.4.3. Mode "P-P"
1. Select

P-P mode by means of the FUNC key.

2. If possible disconnect all low impedance loads downstream the point at which the

measurement is to be taken, as such an impedance would be in parallel with the line
impedance to be measured.

3. Connect the 3 black, green and blue connectors of the single cables in the corresponding

input terminals of the instrument

B1, B3, B4 (see possible connections in the following

pictures). When using untied cables connect the crocodiles to the free ends of the cables.

B2

B3

B4

B1

L1
L2

L3
N

Instrument connection for measurement of phase to phase impedance


4. Connect the crocodiles to the conductors of the three-phase system (see picture).


5. Press

the

START/STOP key. The instrument starts the test.


At the end of the test the instrument

emits a

double sound signal

indicating that the test is correctly
terminated
and displays the values
alongside.

LOOP

1.07

P-P

383

V

374

A


CAUTION

The measurement of Phase-Phase impedance entails the following of a
current of about 11.5A between the above said conductors. This may cause
the tripping of magnetothermal RCDs with nominal value lower than 10A. If
necessary effect the test upstream the RCD itself.

Formula for calculation of prospective short circuit current:

PP

N

CC

Z

U

I

=


where

U

N

= phase to phase voltage 127V if V

meas

≤ 150V

230V if 150V < V

meas

≤ 265V

400V if V

meas

> 265V


SAVING:

The tests can be stored pressing the

SAVE key twice

(according to paragraph 5.1).

Value of phase to
phase line
impedance
expressed in

Ω.

Phase to phase
effective voltage

value expressed
in Volt.

Effective value of the phase to phase prospective
short circuit current expressed in Ampere
calculated according to the following formula.

SAVE

START

STOP