Handbook BM 100 A
1 Agitator
2
Support beam perpendicular to the pulse
direction
3
Abrupt changes in tank cross section
4 Heating
tubes
5
Alternative solution: reference chamber -
electromagnetic field is contained within
chamber.
6
Gauge electromagnetic field :
Any intruding metallic object will be
detected in this zone if perpendicular to
the emitted pulse direction.
= Do not fit the process connection near to these objects.
Probe Type
Recommended minimum distance of probe from
objects inside the tank in millimetres (inches)
Single (types F, H and K)
300 (12)
Twin (types A, B, G and L)
100 (4)
Coaxial (type D)
0 (0)
Electromagnetic field shape around probe, by type (not to scale)
Single (types F, H and K)
Twin (types A, B, G and L)
Coaxial (type D)
No beam angle for any probe type.