Clearing the current gauge volume calibration table data

1.

Go to fct. 1.7.3 DELETE TAB.

2.

Press “right” to enter the data set field. This will read “SURE No”.

3.

Press “up” to select “SURE Yes”.

4.

Press “enter” to confirm clearing the “strapping table”.

5.

Go to fct. 1.7.2 INPUT TAB. to enter new values into an empty table.



3.4.4 Network Configuration – digital outputs and gauge identification : RS 485, multi-drop...

Example procedure 6:
• turn off analog current output to read a digital signal *, **

• give each gauge an address and device number

Step Action

Press buttons to
finish step

Information displayed
at the end of each step

1

Start from Operating Mode screen

n/a

6750
LEVEL mm

2

From operating mode go to fct. 1.3.1
to turn off analog output.*

→,→,↑,↑,→

Fct.

1.3.1

FUNCTION.I.1

3

Enter function ( current value: level)*

LEVEL

4

Set to analog output to “off”*

↑,↑,↑,↑

OFF

5

Exit data set level and go to fct. 1.6.2
to give the gauge an address.

↵,↵,↑,↑,↑,→,↑

Fct.

1.6.2

ADDRESS

6

Enter function. A number will be
displayed (default value: 000- no
network)

000

7

Change address to 001.

→,→,↑

001

8

Enter to confirm and exit to function.
Go to fct 1.4.4 to give the gauge a
device number (displayed on a remote
user interface such as a PC (Personal
Computer) or a HHC (HART Handheld
Controller).

↵,↵,↑,↑,↑,↑,→,↑,↑

Fct.

1.4.4

DEVICE No

9

Enter data set field. (default value:
00000.001). A 9-keystroke
alphanumerical code may be entered
using “up” to scroll through the choices
and “right” to change digit

00000.001

10

New name: 10000.001


10000.001

11

Exit data set field, function, sub menu
and menu. Save configuration and
return to operating mode.

↵,↵,↵,↵

6750

LEVEL

mm

*Ignore steps 2, 3 and 4 for HART® outputs. Leave the analogue output “on”.
**Use local display interface only to configure each BM 100 A meter for network integration.

Communication Rate

Use fct. 1.6.1 BAUD RATE to optimise communication with the remote work station when using
RS485 interface. HART/ SMART networks are set at 1200 baud.