KROHNE BM 100A Handbook User Manual

Page 38

Advertising
background image


38

Handbook BM 100 A


1.6.0 SERIAL I/O

(Serial Input/Output)

For integrating into a digital
network

Select 1200, 2400, 4800,
9600 or 19600 baud

1.6.1 BAUDRATE*

1200 bd

Defines transmission rate of
information between BM100 and
remote terminal. HART/SMART
outputs are fixed at 1200 baud

Enter a number from 0 to
255**

1.6.2 ADDRESS

000

Address of a device in a digital
network. If HART protocol is used,
entering >0 will configure the gauge
for multi-drop networks & output
current 1 to 4 mA.

*RS485 interface only
**0 to 15 if HART.

1.7.0 STRAP. TAB

(Strapping Table)

For calibrating the gauge for
volume measurement.

Select m³, litre, US gal.,
GB, gal., Ft.³, bbl, kg,
metric ton, ton GB or ton
US.

1.7.1 VOL.

UNIT

(Volume / Mass Unit)

Selects the volume or mass unit for a
volume strap table.

Select point 01 to 50,
enter level and then
volume values
respectively.

1.7.2 INPUT.TAB.

(Strapping Table Input)

00

Table or calibrating the gauge point
by point in terms of product level and
volume values.

Select SURE YES or
SURE NO.

1.7.3 DELETE

TAB.

(Strap Table Delete)

SURE NO

Clears the data from the existing
strapping table.