4

Service and maintenance


4.1

Test Functions in the user menu


A series of tests are available in this part of the configuration menu. This permits the local display
and the instrument calibration to be checked.

Test Function

Input range

Description

2.0.0 TEST

2.1.0 TEST DISPL.
(TEST DISPLAY)

Gives full display of all
segments of the LCD display.

2.2.0 CUR OUTP.I
(CURRENT OUTPUT I)

2.2.1

VALUE I 1

N/a

Gives a reading of current from
analogue output 1.

2.2.2

TEST I 1

Select 3.6, 4, 12, 20 or 22 mA.

Forces the output 1 to a
selection of 4 output currents.

2.2.3

VALUE I 2

N/a

Gives a reading of current from
analogue output 2.

2.2.4

TEST I 2

Select 3.6, 4, 12, 20 or 22 mA.

Forces the output 2 to a
selection of 4 output currents.

2.3.0 COMM. TEST

For factory use only.


4.2 Troubleshooting


4.2.1 Parameter

errors


Function

Action to be taken

4 PARAMETER

ERROR

4.1 CURRENT

OUPUT

4.1.1

SCALE I 1 MIN

When a parameter error occurs, the digit 4 is displayed
with the description “PARAMETER ERROR”.

4.1.2

SCALE I 1 MAX

4.1.3

SCALE I 2 MIN

4.1.4

SCALE I 2 MAX

4.2 STRAP

TABLE

4.2.1 STRAP

TABLE

INPUT

4.2.2 STRAP

TABLE

SUPRESS

4.2.3 DISPLAY

MODE

4.2.4 DISPLAY

ITEM

4.2.5 FUNCTION

I

1

4.2.6

FUNCTION I 2

4.3 PARAMETERS

ERRORS

4.3.1 DEAD

ZONE

4.3.2 DETECTION

DELAY


• The RIGHT key should be pressed until the function

concerned is given.


• Press the RIGHT key again to make the value appear

for modification.


• Follow the instructions for resetting the value in

Section 3 and then press ENTER to exit.


• If more than one error is reported, the procedure

should be repeated.