How to configure the gauge to give volume readings
User menu functions 1.71, 1.7.2 & 1.7.3
The BM100 A may be calibrated to measure volume by using a 50-line “strapping (calibration)
table”. The shape of the tank decides the number of points to be entered for accurate measurement.
Do not set User menu function 1.3.1 or 1.3.5 to “Volume” until steps 1 to 4 have been completed.

Example procedure 5
• set volume or weight units for calibration

• program up to 50 points (2 in this example) giving the level and filled volume for each point.

Step Action

Press buttons to
finish step

Information displayed
at the end of each step

1

Start from Operating Mode screen

n/a

6750
LEVEL mm

2

From operating mode go to fct. 1.7.1 to
choose volume units for calibration.

→,→,↑,↑,↑,↑,↑,↑,↑

,→

Fct.

1.7.1

VOL.UNITS

3

Enter function ( default value: m3). A
selection of metric, imperial and federal
weight and volume units are available.

m3

4

Set to “Liter”.

Liter

5

Exit data set level and go to fct. 1.7.2 to
input calibration points.

↵,↑

Fct.

1.7.2

INPUT.TAB

6

Enter function to program the table. The
point is displayed (current value: 15).

15

7

Change to point 01.

↑,↑,↑,↑,↑,↑,↑,↑,↑,

→,↑,↑,↑,↑↑,↑

01

8

Press ENTER to set level (height from the
bottom of the tank)at point 01 in mm.
Current value: 00100.0 mm

00100.0

mm

9

Change level to 00200.0 mm

→→↑

00200.0

mm

9

To confirm and set filled volume at this
level. Default volume: 0003000.0 Liter.

0003000.0

Liter

10

Change to 0000020.0 Liter. Press enter to
confirm and exit data set field.

→,→,→,↑,↑,↑↑,↑,

↑,↑,→,→,↑,↑

0000020.0
Liter

11

To confirm and exit data set field.

Fct.

1.7.2

INPUT.TAB

12

Enter function to add second point on
table. The point number will be displayed
(default value: 15)

15

13

Change to point 02.

↑,↑,↑,↑,↑,↑,↑,↑,↑,

→,↑,↑,↑,↑,↑,↑,↑

02

14

To set level at point 02 in mm. Current
value: 00600.0 mm

00600.0

mm

15

Change level to 00400.0 mm

→,→,↑,↑,↑,↑,↑,↑,↑

,↑

00400.0

mm

16

To confirm and set filled volume at this
point. Current volume: 0799999.9.0 Liter.

0799999.9

Liter

17

Change to 0000040.0 Liter.

→,↑,↑,↑,→,↑,→,↑,

→,↑,→,↑↑,↑,↑,↑,

→,↑,→,↑

0000040.0

Liter

18

Exit data set field, function, sub menu and
menu. Save configuration and return to
operating mode.

↵,↵,↵,↵

6750

LEVEL

mm