Handbook BM 100 A

27


Tank height

Configuration Mode user function 1.1.1

This function is usually either defined as true tank height or as factory configured probe length,
L

1

if the former is not supplied by the customer in the order.



Why change the tank height?

• setting the parameter in Function 1.1.1 to L

1

avoids having a non-measurable zone

underneath the probe where the measurement on the display freezes.

• when setting up a measurement scale as explained on the following pages, this means that

the level at the end of the probe will be taken as zero instead of the tank bottom.


How tank height affects measurement when either Level or Distance is measured

1

True tank height

2 Measurable

height (factory
configured probe
length, L

1

)

3 Non-measurable

zone

4

With true tank
height (1) set in
Function 1.1.1 of
the User’s menu.

5 With

factory-

configured probe
length, L

1

, set in

Function 1.1.1. of
the User’s menu.



Example procedure 1:
• to change true tank height (10000mm) to factory configured probe length, L

1

(9000mm) , and

then save the new parameter. Refer to item 5 in the diagram above.


Step

Action

Press buttons to
finish step

Information displayed at the end of
each step

1

Start from Operating
Mode screen

n/a

6750
LEVEL mm

2

Go to function (fct.) 1.1.1
from Operating Mode

→,→,→

Fct.

1.1.1

TANK

HEIGHT

3

Enter function (current
value: 10000mm)

10000.00

4

Modify value to 09000.00
(millimetres)

↑,↑,↑,↑,↑,↑,↑,↑,↑,

→,↑,↑,↑,↑,↑,↑,↑,↑,

09000.00

5

Exit to Save function

↵,↵,↵,↵

STORE

Yes

6

Save & exit to Operating
Mode

5750
LEVEL mm