4.2.2 Basic Servicing – Replacing fuses and the electronics chassis


BM100 servicing by the customer is limited by warranty to the removal and
replacement of :
• the power supply fuse

• the electronics chassis.

Other repairs must be done by KROHNE-authorised service staff.


Removing the electronics chassis

Before changing the electronics chassis remember to save the current configuration
of the instrument :
• Note the settings on the configuration record supplied in appendix B or

• Save the settings in a *.dat file on PC-STAR. Press the F4 key to open a

directory.

Please follow these instructions carefully:

Step 1
Disconnect instrument from the power supply. If instrument is equipment category 1 /
2 GD (ATEX classification) or located in a Division 1 (US) area, please refer to
Supplementary Installation and Operating Instructions BM100 A / Ai KEMA 01 ATEX
1078X for more information.


Step 2
Remove back cover with supplied wrench.

Step 3

Unplug terminal blocks from electronics chassis.

Step 4
Remove the front cover with wrench.

Step 5

Undo the 2 screws holding display on chassis.