Handbook BM 100 A
3.4.3 Quick
Configuration: configuration examples
The minimum functions (fct.) to be configured for a simple measurement are listed below:
• 1.1.1
Tank Height
• 1.2.1-6 Display Functions
• 1.4.2-3 Entry Code 1 / Code 1
• 1.3.1-4 Current Output 1 (& 2)
• 1.7.1-2 Volume calibration*
*For volume measurements
Example procedures for each set of functions are given on the following pages. Each procedure is
given in a series of steps in table form and begins from the Operating Mode.
Useful definitions for quick configuration
Typical gauge used for quick configuration examples
Probe type:
twin Ø4mm/0.16” cable probe, type L
Tank height (Fct. 1.1.1):
10000.00mm/33ft
Hold distance (Fct. 1.1.2)
0.25m/10ft (see “probe measurement limits” in section 6.2.3 for
the Ø4mm twin cable probe, type L)
Probe length, L
1
(Fct. 1.1.7):
9.00m/29.5ft (Do not modify unless advised to)