Measurement range setting (Function mode 1)

4.6.2

This function is used to set the measurement range on internal oscillation emission. The
measurement range has two levels. The indicated contents depend on the measurement unit
setting.

Measurement unit

*Measurement range

fpm

60 to 12,000

(Available setting range: 60.0 to 12,000.0)

60 to 120,000

(Available setting range: 60.0 to 12,000.0)

* The emission count saved in the memory will be x10 or 1/10 by changing the measurement range.

Example: Save the emission count (frequency) of 6000 fpm in the range between 60 fpm and 120,000

fpm. After that, the saved value of the emission count (frequency) will be 600.0 fpm by
changing the measurement range to between 60 fpm and 12,000 fpm.

* If the emission count (frequency) setting value is set to the one beyond (or below) the range when

changing the measurement range, the value will be limited to the maximum (or minimum) value within
the range.
Example 1: Save the emission count (frequency) of 60,000 fpm in the range between 60 fpm and

120,000 fpm. After that, the saved value of the emission count (frequency) will be 12,000.0
fpm by changing the measurement range to between 60 fpm and 12,000 fpm.

Example 2: Save the emission count (frequency) of 100 fpm in the range between 60 fpm and 120,000

fpm. After that, the saved value of the emission count (frequency) will be 60.0 fpm by
changing the measurement range to between 60 fpm and 12,000 fpm.

Turn the dial to the CW direction to set the range to between 60 fpm and 120,000 fpm, and to the
CCW direction to set the range to between 60 and 12,000.
Operate the dial to stop the alternate display, then only the measurement range setting value is
indicated.

Operation

Display

Change the setting

value

Indicates alternately

* Previous setting value

CCW

CW

Go to function mode

2

Go to function mode 2