Instruction Manual PC 510

10

Calibrating for Conductivity:

1. If necessary, press the MODE key to select conductivity mode.

2. Rinse the probe thoroughly with de-ionised water or a rinse

solution, then rinse with a small amount of calibration standard.

3. If necessary, ensure that the probe’s yellow probe guard is

attached. Dip the probe into the calibration standard. Immerse the
probe tip beyond the upper steel band. Stir the probe gently to
create a homogeneous sample.

4. Wait for the measured conductivity value to stabilise. If the READY

indicator has been activated (SETUP program P7.1 – see page
31), the READY annunciator lights when the reading is stable.

5. Press CAL/MEAS to enter conductivity or TDS calibration mode.

The CAL indicator will appear in the upper right corner of the
display.

6. Press

the

or key to change the value on the primary display

to match the value of the calibration standard.

7. Press ENTER to confirm calibration value. The meter returns to the

MEAS (measurement) mode.

8. Repeat steps 1 to 7 for other measuring ranges.


NOTES:
When entering calibration mode, the meter will display the
factory default value. If the meter was previously calibrated, the display
may “jump” to the factory default / uncalibrated value when switching
from measurement to calibration mode.

To exit from Conductivity calibration mode confirming calibration,
DO NOT press the ENTER key in step 7. Press CAL/MEAS instead.

This will retain the meter’s old calibration data in the measuring range
of the calibration.

You can offset the conductivity reading up to ±40% from default
setting. If your measured value differs by more than ±40% clean or
replace probe as needed, or use a calibration standard with a higher
value as required.

A wide selection of high-quality calibration standards is available. See page 44 for more information.