Instruction Manual PC 510

25

8.2

P1.0: Viewing previous pH calibration data

This mode lets you recall previous pH calibration data, which helps you
know when to re-calibrate your meter. This is a “view-only” mode.

From measurement mode:

1. Press the MODE key to select pH measurement mode if necessary.

2. Press the SETUP key to enter Setup mode.

3. Press

the

or key to scroll through sub-groups until you view

parameter P1.0.

4. Press the ENTER key repeatedly to view previous calibration data.

See Figure 24.

USA

NIST

• P1.1 = pH 1.68

1.68

• P1.2 = pH 4.01

4.01

• P1.3 = pH 7.00

6.86

• P1.4 = pH 10.01

9.18

• P1.5 = pH 12.45

12.45

5. When you have scrolled through all calibration data, you will

automatically return to the sub-group menu. Press CAL/MEAS key
to return to measurement mode.


NOTES: If there is no previous calibration data at a particular point, the
primary display will show “----“.

Figure 25: P1.0 - View

calibration data for pH