P 1 .1, P 1 .3, P 1 .0 – Thermo Fisher Scientific CyberScan PC 510 User Manual
Page 29: P 1 .4

Instruction Manual PC 510
25
8.2
P1.0: Viewing previous pH calibration data
This mode lets you recall previous pH calibration data, which helps you
know when to re-calibrate your meter. This is a “view-only” mode.
From measurement mode:
1. Press the MODE key to select pH measurement mode if necessary.
2. Press the SETUP key to enter Setup mode.
3. Press
the
or key to scroll through sub-groups until you view
parameter P1.0.
4. Press the ENTER key repeatedly to view previous calibration data.
See Figure 24.
USA
NIST
• P1.1 = pH 1.68
1.68
• P1.2 = pH 4.01
4.01
• P1.3 = pH 7.00
6.86
• P1.4 = pH 10.01
9.18
• P1.5 = pH 12.45
12.45
5. When you have scrolled through all calibration data, you will
automatically return to the sub-group menu. Press CAL/MEAS key
to return to measurement mode.
NOTES: If there is no previous calibration data at a particular point, the
primary display will show “----“.