P 2.0, P2.1, P 2.2 – Thermo Fisher Scientific CyberScan PC 510 User Manual

Page 30

Advertising
background image

Instruction Manual PC 510

26

8.3

P2.0: Viewing pH electrode data

Program 2 has two “view-only” options that let you check the electrode
parameters for diagnostic purposes. It lets you view:

• P2.1 = Electrode offset

• P2.2 = Electrode slope

From pH measurement mode

1. Press the MODE key to select pH measurement mode.

2. Press the SETUP key to enter Setup mode.

3. Press

the

or key to scroll through sub-groups until you view

parameter P2.0.

4. Press the ENTER key to select parameter 2.1.

5. The display shows the electrode offset value. It is the mV offset at

pH 7.00. If you have not calibrated at any buffer, the primary display
shows 0.00 mV.

6. Press the ENTER key to proceed to P2.2.

7. The display shows electrode slope in percentage. Slope displayed

is the average slope based on the pH calibrations. Default setting is
100.0.

8. At any point, you can press the CAL/MEAS key to return to

measurement mode.

Figure 26: Viewing electrode's

offset and slope status from

pH measurement mode

SETUP

p 2.0

ele

pH

SETUP

5.3

p2.1

mV

SETUP

p 2.2

98.7

%

ENTER

RANGE

ENTER

RANGE