Instruction Manual PC 510

31

8.5

P4.0: Resetting to factory default settings (pH)

This program lets you reset all pH parameters to factory default settings. This clears all calibration
data and any other pH setup functions you might have changed.

The following settings will remain as you have set them:

• Temperature unit of measure (°C or °F)

• The temperature offset calibration value

• All conductivity calibration data and parameters

From measurement mode

1. Press the MODE key to select pH measurement mode.

2. Press SETUP key to enter Setup mode.

3. Press

the

or key to scroll through sub-groups until you view

parameter P4.0.

4. Press the ENTER key to enter parameter P4.0. See Figure 32.

5. Press

the

or to toggle between NO and YES.

• NO retains current settings

• YES resets to factory default settings

6. Press the ENTER key to confirm selection and to return to the measurement mode. Otherwise

press CAL/MEAS key to return to measurement mode without resetting to factory default.

NOTE:

To clear all conductivity & TDS data, see page 38.

See Addendum 4 on page 49 for a table of factory default settings.

Figure 32: P4.0 - Reset to

factory default values

SETUP

MEAS

SETUP

MEAS

yes

pH

pH