P 5.0, P 5.1, P 5.3 – Thermo Fisher Scientific CyberScan PC 510 User Manual

Page 36: P 5.4

Advertising
background image

Instruction Manual PC 510

32

8.6

P5.0: Viewing previous conductivity calibration data

This mode lets you recall previous calibration data, which helps you know when to re-calibrate your
meter. This is a “view-only” mode.

This function applies for conductivity & TDS mode.

From conductivity or TDS mode

1. Press the MODE key to select conductivity or TDS measurement

mode.

2. Press the SETUP key to enter Setup mode.

3. Press

the

or key to scroll through sub-groups until you view

parameter P5.0.

4. Press ENTER key repeatedly to view previous calibration data.

5. Each calibration data corresponds to each measurement range.

• P5.1 = Range 1 (0.00 – 19.99 µS or 0.0 – 9.99 ppm)

• P5.2 = Range 2 (0.0 – 199.9 µS or 0.0 – 99.9 ppm)

• P5.3 = Range 3 (0 – 1999 µS or 0 – 999 ppm)

• P5.4 = Range 4 (0.00 – 19.99 mS or 0.00 – 9.99 ppt)

• P5.5 = Range 5 (0.0 – 199.9 mS or 0.0 – 99.9 ppt)

6. When you have scrolled through all calibration data, you will

automatically return to the sub-group menu. Press CAL/MEAS key
to return to measurement mode.

NOTE: If there is no previous calibration data at a particular point, the
primary display will show “----“.

Figure 33: View calibration

data in conductivity probe

SETUP

cal

p 5.0

mS

S

ppt
ppm

µ

SETUP

---

p 5.1

ENTER

RANGE

SETUP

CAL

1413

p 5.3

µ

S

ENTER

RANGE

SETUP

CAL

1

2.88

p 5.4

mS

ENTER

RANGE