Instruction Manual PC 510
20
7 HOLD
FUNCTION
This feature lets you freeze the display for a delayed observation. HOLD can be used any time in
MEAS mode.
1. To hold a measurement, press the HOLD key while in measurement mode. “HOLD” will appear on
the display.
2. To release the held value, press the HOLD key again. Continue to take measurements
MEAS
HOLD
READY
ATC
°C
678
20.7
µS
NOTE:
• If the meter is shut off manually, the HOLD value will be lost.
• Your meter has an auto endpoint feature. When this feature is switched on, and when a
reading is stable for more than 5 seconds, the display will automatically “HOLD” the
reading. The “HOLD” indicator appears. Press the HOLD key to release the reading. To
switch on or off the auto endpoint feature, see SETUP Program P3.1 on page 28 for pH
and SETUP Program P7.1 on page 34 for conductivity/TDS.
8 ADVANCED
SETUP
FUNCTIONS
The advanced setup mode lets you customise your meter’s preferences and defaults. The PC 510
bench meter features different sub-groups that organise setup parameters.
The sub-groups are:
For pH
For Conductivity & TDS
P1.0
View calibration data (CAL)
P5.0
View calibration data (CAL)
P2.0
View electrode data (ELE)
P6.0
View electrode data (ELE)
P3.0
Unit Configuration (COF)
P7.0
Unit Configuration (COF)
P4.0 Reset
to
factory
default
settings (rSt)
P8.0 Temperature
(tPr)
P9.0
Reset to factory default settings
(rSt)
Figure 23: HOLD function