Thermo Fisher Scientific CyberScan PC 510 User Manual

Page 18

Advertising
background image

Instruction Manual PC 510

14

6 MEASUREMENT

6.1

Taking pH Measurements

6.1.1 Automatic

Temperature

Compensation

Automatic Temperature Compensation only occurs when a temperature sensing element is plugged
into the meter.

Temperature sensing element refers to the following probes made specifically for this meter. See
accessories for ordering information

The conductivity electrode with a built-in temperature

sensor

Temperature probe; or

The 3-in-1 pH/Temperature combination electrode.

If there is no temperature sensor plugged into the meter, the default
manual temperature setting is automatically 25

°C. You can manually

set the temperature to match your working conditions using a separate
thermometer.

For automatic temperature compensation (ATC) simply plug the temperature probe into the meter (see
page 5 for directions). The ATC indicator will light up on the LCD. See Figure 13.

NOTE: The temperature sensing element must be submersed in the liquid you are measuring.

6.1.2

Manual Temperature Compensation (pH)

IMPORTANT: For manual compensation, you must disconnect the
temperature probe (see page 5 for instructions).

1. Switch the meter on. Press MODE key to select pH mode.

2. Press the CAL/MEAS key to enter pH calibration mode. The CAL

indicator will appear above the primary display.

3. While in pH calibration mode, press the MODE key to enter

temperature calibration mode. The primary and secondary displays
show the last set temperature value.

4. Check the temperature of your sample using an accurate

thermometer.

5. Press

the

or

key to set the temperature to the measured

value from step 4.

6. Press ENTER to confirm the selected temperature and to return to

the pH measurement mode.

See Figure 14.

The meter will now compensate pH readings for the manually set
temperature.

NOTES: To exit this program without confirming the manual
temperature compensation value, DO NOT press ENTER in step 6.
Press CAL/MEAS instead.

Figure 13: ATC annunciator will

light up when connected to

temperature probe

Figure 14: Manual temperature

compensation

MEAS

ATC

pH

°C

7.05

24.5

READY

CAL

°C

25.0

25.0

MEAS

°C

7.05

30.0

READY

pH

ENTER

RANGE

°C

30.0

25.0

CAL