Instruction Manual PC 510

50

18 ADDENDUM 5: METER FACTORY DEFAULT SETTINGS

Resetting the meter to factory default settings clears all calibration data and most other setup
functions you might have changed. The following settings will remain as you have set them:

• Temperature unit of measure (°C or °F)

• The temperature offset calibration value.

NOTE: Conductivity and pH data are cleared separately from each other.

• To clear pH data, see Sub group P4.0.

• To clear conductivity/TDS data, see Sub group P9.0.

Type Parameter

Default

Remarks

pH parameters

P1.1
P1.2
P1.3
P1.4
P1.5

View pH calibration data

----
----
----
----
----

No calibration data for 1

st

buffer, pH 1.68

No calibration data for 2

nd

buffer, pH 4.01

No calibration data for 3

rd

buffer, pH 7.00 or 6.86

No calibration data for 4

th

buffer, pH 10.01 or 9.18

No calibration data for 5

th

buffer, pH 12.45

P2.1
P2.2

View electrode offset
View electrode slope

0.00 mV
100.0%

No offset adjustment
No slope adjustment

P3.1
P3.2
P3.3
P3.4

Ready indicator
# pH calibration points
Buffer set option
°C or °F

Ready on
3
USA
No default

Ready indicator on; auto endpoint off
3 pH calibration points available (1 – 5 pt range)
USA or NIST buffer set
°C or °F remains as selected

P4.0

pH factory default

No

Retains your current settings

Conductivity and TDS parameters

P5.1
P5.2
P5.3
P5.4
P5.5

Viewing conductivity or TDS
calibration data

----
----
----
----
----

No calibration data for range 1
No calibration data for range 2
No calibration data for range 3
No calibration data for range 4
No calibration data for range 5

P6.1
P6.2
P6.3
P6.4
P6.5

Viewing conductivity or TDS
calibration data

1.0
1.0
1.0
1.0
1.0

No offset for effective cell constant for range 1
No offset for effective cell constant for range 2
No offset for effective cell constant for range 3
No offset for effective cell constant for range 4
No offset for effective cell constant for range 5

P7.1
P7.2
P7.3
P7.4

Ready indicator
°C or °F
ATC on or off
TDS factor

Ready on
No default
ATC on
0.5

Ready indicator on; auto endpoint off
°C or °F remains as selected
----
Adjustable from 0.4 to 1.0

P8.1
P8.2

Temperature coefficient
Normalisation temperature

2.1% per

°C

25

°C

Adjustable from 0 to 10%
Adjustable from 15 to 30

°C

P9.0

Conductivity factory default

No

Retains your current settings