Thermo Fisher Scientific CyberScan PC 510 User Manual

Page 23

Advertising
background image

Instruction Manual PC 510

19

Manual-ranging

The manual ranging function lets you select the specific range you want to work in. Refer to page 9 for
the table of range.

1. To select the desired measuring range manually, press the RANGE

key while in measurement mode. The meter will lock on to the
appropriate range and the “MEAS” indicator blinks.

2. Press RANGE key again (if needed) until desired range is selected.

3. To re-select the Auto-ranging function, repeatedly press the RANGE

key until the “MEAS” indicator appears without blinking.


NOTES:

The meter will not let you manually select a range in which the reading will be over-range.

The meter resets to the Auto-ranging function once it is turned off. You will have to reset the manual
ranging function each time you turn the meter off.