Instruction Manual PC 510

34

8.8

P7.0: Conductivity or TDS measurement configuration

8.8.1

P7.1: READY indicator and auto endpoint function

This program lets you select:

“READY indicator on” to indicate when the reading is stable.

“READY indicator off” for faster meter response.

Auto endpoint function on. Select auto endpoint on to “hold” the

reading when it is stable for more than 5 seconds. The display
automatically freezes, and the HOLD indicator appears on the left
side of the display. Press the HOLD key to release the display and
access other functions.

From measurement mode

1. Press the MODE key to select conductivity or TDS measurement

mode.

2. Press the SETUP key to enter Setup mode.

3. Press

the

or key to scroll through sub-groups until you view

parameter P7.0.

4. Press ENTER key to select parameter P7.1.

5. Press

the

or key to select the configuration you require.

• OFF switches the READY indicator off.

• ON switches the READY indicator on.

• ON and HOLD together switches the auto endpoint feature on.

6. Press ENTER key to confirm selection and to proceed to step 4 of

P7.2. You can also press the CAL/MEAS key to return back to
measurement mode.

NOTE: Meter default is set for Ready indicator on, and auto endpoint function off.