Thermo Fisher Scientific CyberScan PC 510 User Manual

Page 17

Advertising
background image

Instruction Manual PC 510

13

5.7 Temperature

Calibration

The conductivity electrode (EC-CONSEN91W / 35608-50) supplied has a
built-in temperature sensor. Alternatively, a separate temperature sensing
element can be used (such as temperature probe EC-WPPHTEM-01W /
35618-05), or a “3-in-1” pH/Temperature combination electrode with ATC
connection.

The conductivity probe is factory calibrated. Temperature calibration is
recommended only if you suspect temperature errors may have occurred
over a long period of time, or if you have a replacement probe.

Temperature calibration is accessible during pH, conductivity or TDS
calibration.

Temperature Calibration

1. Make sure the conductivity electrode, or temperature probe, or “3-in-1”

electrode is attached to the 6-pin connector.

2. Switch the meter on.

3. Press the CAL/MEAS key to enter calibration mode (either from pH or

conductivity mode). The CAL indicator will appear above the primary
display.

4. While in pH (or conductivity or TDS) calibration mode, press the MODE

key to enter temperature calibration mode. The primary display shows
the last set temperature value and the secondary display shows the
temperature reading with zero offset.

5. Dip the ATC probe into a solution of known temperature (i.e. a

temperature bath). Allow time for the temperature probe to stabilise.

6. Scroll with the

or key to set the correct temperature value (i.e. the

temperature of the temperature bath). You can adjust the reading by 0.1
°C or °F increments.

7. Once you have selected the correct temperature press the ENTER key.

The meter automatically returns to measurement mode.

NOTES:

• You can offset the temperature reading up to ±5° from default reading.

• To exit this program without confirming the temperature calibration value, DO NOT press ENTER,

press CAL/MEAS in step 7 instead.

Figure 12: Temperature calibration

in pH mode

MEAS

pH

ATC

°C

7.1

6

22.3

MEAS

pH

ATC

°C

7.1

6

22.0

CAL

pH

pH

7.1

6

7.00

CAL

ATC

°C

22.3

22.3

ATC

°C

22.0

22.3

CAL

READY

MODE

ENTER

RANGE

CAL

MEAS