Instruction Manual PC 510
22
8.1
Advanced SETUP mode Overview
1. In either pH or Conductivity measurement mode, press the SETUP key to enter Setup mode.
2. Press
the
or key to scroll through sub-groups.
3. Press ENTER key to enter a particular parameter.
See Addendum 4 on page 49 for a table of meter factory default settings.
SETUP in pH measurement mode
P1.0: View previous calibration data
P1.1
First calibration point (pH 1.68)
P1.2
Second calibration point (pH 4.01)
P1.3
Third calibration point (pH 7.00 or 6.86)
P1.4
Fourth calibration point (pH 10.01 or 9.18)
P1.5
Fifth calibration point (pH 12.45)
P2.0: View electrode data
P2.1: pH
electrode
offset
P2.2: pH
electrode
slope
P3.0: Unit Configuration
P3.1
READY indicator and auto endpoint function – select on or off
P3.2
Number of pH calibration points: 2, 3, 4 or 5
P3.3
Select USA or NIST buffer set
P3.4 Select
°C or °F
P4.0: Reset to factory defaults
P4.0
Reset meter to factory defaults