Instruction Manual PC 510

28

8.4.1

P3.1: READY Indicator and auto endpoint function

This program lets you select:

“READY indicator on” to indicate when the reading is stable

• “READY indicator off” for faster meter response.

Auto endpoint function on. Select auto endpoint on to “hold” the

reading when it is stable for more than 5 seconds. The display
automatically freezes, and the HOLD indicator appears on the left
side of the display. Press the HOLD key to release the display and
access other functions.

From measurement mode

1. Press the MODE key to select pH measurement mode.

2. Press SETUP key to enter Setup mode.

3. Press

the

or key to scroll through sub-groups until you view

parameter P3.0.

4. Press the ENTER key to select parameter 3.1.

5. Press

the

or key to select the configuration you require.

• OFF switches the READY indicator off.

• ON switches the READY indicator on.

• ON and HOLD together switches the auto endpoint feature on.

6. Press the ENTER key to confirm selection and to proceed to step 4

of P 3.2. Press the CAL/MEAS key to return to measurement mode.


NOTES: Meter default is set for READY indicator on and auto endpoint function off.

You can also change the READY indicator and auto endpoint function in Program P7.1 (available from
conductivity or TDS mode). Any changes you make to the READY indicator/auto endpoint function in
pH mode will also change in conductivity mode.

Figure 28: P3.1 - Selecting

READY function

SETUP

READY

HOLD
ON

p 3.1

SETUP

READY

OFF

p 3.1

SETUP

READY

ON

p 3.1

SETUP

cof

p 3.0

ENTER

RANGE