Vaisala DM500 User Manual

Page 105

Advertising
background image

Chapter 16 _____________________________________________________ Serial Line Commands

VAISALA ______________________________________________________________________ 103

The CDP* command displays also the coefficients of the correction
polynomial, as well as the reference value and residual error for each
calibration data point.

Example:

>cdp<cr>
1. reference : -39.99 <cr>
1. measurement : -40.009 <cr>
2. reference : -30.01 <cr>
2. measurement : -30.030 <cr>
3. reference : -20.01 <cr>
3. measurement : -20.026 <cr>
4. reference : -9.98 <cr>
4. measurement : -9.977 <cr>
5. reference : 0 <cr>
5. measurement : -0.006 <cr>
6. reference : 10 <cr>
6. measurement : 10.002 <cr>
7. reference : 19.99 <cr>
7. measurement : 19.982 <cr>
8. reference : 29.98 <cr>
8. measurement : 29.989 <cr>
9. reference : 39.98 <cr>
9. measurement : 39.989 <cr>
10. reference : <cr>
<cr><lf>
Reference Measurement<cr><lf>
-39.990 -40.009<cr><lf>
-30.010 -30.030<cr><lf>
-20.010 -20.026<cr><lf>
-9.980 -9.977<cr><lf>
0.000 -0.006<cr><lf>
10.000 10.002<cr><lf>
19.990 19.982<cr><lf>
29.980 29.989<cr><lf>
39.980 39.989<cr><lf>
Overwrite previous calibration ? Y/N<cr><lf>
OK<cr><lf>
>

>cdp *<cr>
1. reference : -39.99 <cr>
1. measurement : -40.009 <cr>
2. reference : -30.01 <cr>
2. measurement : -30.030 <cr>
3. reference : -20.01 <cr>
3. measurement : -20.026 <cr>
4. reference : -9.98 <cr>
4. measurement : -9.977 <cr>
5. reference : 0 <cr>
5. measurement : -0.006 <cr>
6. reference : 10 <cr>
6. measurement : 10.002 <cr>
7. reference : 19.99 <cr>

Advertising