User's Guide ______________________________________________________________________

68 __________________________________________________________________ M210205EN-D

The serial output command R shows frozen values. You can cancel
the testing sequence by pressing the

CANCEL

.

6.

When the test is completed, you see the following display:

7.

Press

EXIT

to return to the main view.

The following factors affect the cooling capacity:

- Temperature of the environment
- Temperature of the cooling water and
- Gas flow rate of the sample.
Please, check that these settings are correct, see chapter Setting up for
Operation
, page 35. The graphs on pages 27 and 29 show the lowest
achievable sensor temperatures. In addition, to obtain correct results,
the test should be started with the system at equilibrium with room
temperature.

Error History

The error history shows on the display six last shown error messages.

1.

Open

MENU

: press .

2.

Select ►

Measuring

, press .

3.

Select ►

Self diagnostics

, press .

4.

Select

Error history

.

5.

Press

SHOW

to see the last shown error display. Press

NEXT

to

see the earlier ones.

6.

To clear the error history, press

CLEAR

.

7.

Press

EXIT

to return to the main view.