Operational commands – Vaisala DM500 User Manual

Page 93

Advertising
background image

Chapter 16 _____________________________________________________ Serial Line Commands

VAISALA _______________________________________________________________________ 91

3.

Output needed: Relative humidity and pressure to be shown with
a unit. The fields are separated with " " marks. Carriage return
and line feed in the end.

>form 2.1 rh " " u " " 2.4 p " " u \r \n
2.1 RH " " U " " 2.4 P " " U \r \n
>send
23.0 % 1.0164 bar

Operational Commands

START Starting the Measurement

Syntax:

START<cr>

Starts the dewpoint measurement and activates the cooling system.

STOP Stopping the Measurement

Syntax:

STOP<cr>

Stops the dewpoint measurement and disables the cooling system.

PUMP Starting/Stopping the Optional Sample Gas
Pump

Syntax:

PUMP ON/OFF<cr>

Example:

>pump
Pump : OFF
>
>pump on
Pump : ON
>
>pump off
Pump : OFF
>