CAUTION

The dip switches behind the DMI500 MUST NOT be touched,
because the processor may be destroyed as a result of false dip switch
positions.

NOTE

Before connecting a PC to DMI500 or DMP501, make sure there is
no ground potential difference between PC and the DM500.

Serial Communication Settings

Connect the DM500 to PC by using a serial cable (provided). When
connecting EIA-232 (RS232), use the port called 'SERIAL'. When
connecting optional RS485, use the port called 'RS485/LON'. Connect
the RS485 as follows:

Table 2

4-Wire Connection

Terminal

Data Line

1

R+ (input to DM500) ←

2

R- (input to DM500) ←

3

T+ (output from DM500) →

4

T- (output from DM500) →

5

Shield

Table 3

2-Wire Connection

Terminal

Data Line

1

R/T+ connect the same line to
terminals 1 and 3

2

R/T- connect the same line to
terminals 2 and 4

3

R/T+ connect the same line to
terminals 1 and 3

4

R/T- connect the same line to
terminals 2 and 4

5

Shield

Select the DM500 communication mode and data transfer settings as
follows:

1.

Open

MENU

: press .

2.

Select ►

Interfaces

, press .