User's Guide ______________________________________________________________________

112 _________________________________________________________________ M210205EN-D

Displayed Quantities

When downloading recorded data or using the real-time window, the
quantities and units are determined by the current selections in the
DM500's "Quantities and units" menu at downloading time.

Measurement uncertainty values are shown in own data table columns
after quantities Td/f, Td and Pw (if some of these quantities are
active).

Sensing Unit Status

The status information of the sensing unit is shown in numeric form in
"Status" data table column. The numeric codes are:

-

measuring is stopped (STOP)

0

measuring is stabilizing (WAIT)

1

measuring has been stabilized (OK)

-1

error in sensing unit (ERR)

Real-Time Quantities

You can activate only three quantities with "Real-time settings" form.
If you need to see four quantities in the device display and real-time
window, select the quantities and units with DM500 console.

Real-Time Update Interval

DM500 saves real-time data with minimum interval of 15 seconds. If
you choose a shorter real-time update interval in "Real-time settings",
you will receive each value several times and the true time resolution
will not be better.

Chart

You can draw up to three data series simultaneously onto graph.
Measurement uncertainty data is drawn as an independent quantity.