Swagelok tube fittings' installation instructions – Vaisala DM500 User Manual

Page 38

Advertising
background image

User's Guide ______________________________________________________________________

36 __________________________________________________________________ M210205EN-D

Swagelok tube fittings' installation
instructions

1.

Insert the tubing into the Swagelok tube
fitting. Tubing should rest firmly on the
shoulder of the fitting. The nut should be
finger tight.

2.

Mark the nut at the 6 o'clock position.

3.

Hold the fitting body with a backup wrench
and tighten the nut 1¼ turns. Watch the
marking and make one complete turn and
continue to 9 o'clock position.

CAUTION

The DM500 as delivered from the factory has 6 mm Swagelok tube
fittings. If you use ¼" tubing, always use the proper ¼" adapter
fitting (provided). Make sure the adapter is connected the right way.
An incorrectly connected adapter or connecting ¼" tubes straight into
the 6 mm tube fittings may lead to product damage. See Figure 12
and Figure 13 for visual instructions.

Figure 12

Standard 6 mm Swagelok Tube Fitting with the
Elbow Connector in the DMP501 Back Panel

Figure 12 shows the standard set-up of the DMP501 back panel.
Figure 13 illustrates the correct way of connecting ¼" tubing to the 6
mm tube fitting.

Advertising