7. measurement : 19.982 <cr>
8. reference : 29.98 <cr>
8. measurement : 29.989 <cr>
9. reference : 39.98 <cr>
9. measurement : 39.989 <cr>
10. reference : <cr>
<cr><lf>
Reference Measurement<cr><lf>
-39.990 -40.009<cr><lf>
-30.010 -30.030<cr><lf>
-20.010 -20.026<cr><lf>
-9.980 -9.977<cr><lf>
0.000 -0.006<cr><lf>
10.000 10.002<cr><lf>
19.990 19.982<cr><lf>
29.980 29.989<cr><lf>
39.980 39.989<cr><lf>
Overwrite previous calibration ? Y/N<cr><lf>
2.40135193E-01<cr><lf>
-2.59960938E+00<cr><lf>
9.08935547E+00<cr><lf>
-1.30498052E+01<cr><lf>
-4.00089979E+01 -1.94864569E-02<cr><lf>
-3.00300002E+01 -1.91654568E-02<cr><lf>
-2.00259995E+01 -1.26176178E-02<cr><lf>
-9.97700024E+00 -5.47681761E-03<cr><lf>
-6.00000000E-03 -1.04680657E-03<cr><lf>
1.00020003E+01 -2.92509794E-05<cr><lf>
1.99820004E+01 -7.64131546E-04<cr><lf>
2.99889994E+01 8.67143250E-04<cr><lf>
3.99889994E+01 1.14340186E-02<cr><lf>
OK<cr><lf>
>

CTP Adjusting the Dewpoint Temperature (in two
points)

Syntax:

CTP<cr>

1.

Turn the switch 4 (see Figure 4 on page 18, service switches) to
ON-position to enable the adjustment. After adjustment return
the switch 4 to OFF-position.

2.

Set the temperature sensor to the reference conditions. Wait
until the reading is stabilized.

3.

Type the command CTP and give the value of the first
reference, press ENTER.

CTP<cr>

>ctp
Tp : 20.0000 1. ref ? 21
Press any key when ready ...