Vaisala DM500 User Manual
Page 94
Advertising
User's Guide ______________________________________________________________________
92 __________________________________________________________________ M210205EN-D
HEAT Starting/Stopping the Optional Heater
Syntax:
HEAT ON/OFF<cr>
Example:
>heat
Heat : OFF
>
>heat on
Heat : ON
>
>heat off
Heat : OFF
>
SALT Enables/Disables the Salt Contamination
Detection
Syntax:
SALT ON/OFF [interval]<cr>
Default: SALT ON
Example:
>salt
Salt sensing : OFF 24
>
Salt sensing is off.
>salt on
Salt sensing : ON 24
>
Salt sensing is on and detection interval is 24 h.
>salt 30
Salt sensing : ON 30 <cr><lf>
>
Salt sensing is on and detection interval is 30 h.
>salt on 24
Salt sensing : ON 24
>
Salt sensing is on and detection interval is 24 h.
Advertising