Chapter 5 ________________________________________________ Information On Optional Parts

VAISALA _______________________________________________________________________ 31

CHAPTER 5

INFORMATION ON OPTIONAL PARTS

This chapter provides you with information on the optional parts of
the DM500.

The following optional parts are
available for the DM500:

- Heated gas tubes (internal)
- External temperature sensor
- Sampling pump
- Pressure sensor
- Flowmeter

DEWCAP
dewpoint sensor

gas inlet
high pressure

gas outlet

DMP501 internal ONV

(output needle valve)

adjusted from

front panel

optional internal

flow meter

optional pump

optional pressure sensor

Optional Heated Gas Tubes

Heated gas tubes shall be used if the measured dewpoint is high, close
or above ambient temperature. It is recommended to use heated gas
tubes already for dewpoints 5 ºC (9 ºF) below ambient temperature.

Internal gas tubes

The optional heated gas tubes keep the internal gas tubes and sensor
chamber temperature above 60ºC.