17 wind direction sensor orientation, 1 determining true north and sensor orientation, Wind direction sensor orientation – Campbell Scientific UT10 10 foot Tower w/ Grounding Kit and Hinged Base User Manual

Page 57: Determining true north and sensor orientation

Advertising
background image

UT10 Weather Station

8.5.17 Wind Direction Sensor Orientation

8.5.17.1 Determining True North and Sensor Orientation

Orientation of the wind direction sensor is done after the datalogger has been

programmed, and the location of True North has been determined. True North is

usually found by reading a magnetic compass and applying the correction for

magnetic declination*; where magnetic declination is the number of degrees

between True North and Magnetic North. Magnetic declination for a specific site

can be obtained from a USFA map, local airport, or through the National

Geophysical Data Web site at:

www.ngdc.noaa.gov/geomag

. A general map

showing magnetic declination for the contiguous United States is shown in

FIGURE 8-31.

Declination angles east of True North are considered negative, and are subtracted

from 360 degrees to get True North as shown FIGURE 8-32 (0° and 360° are the

same point on a compass). Declination angles west of True North are considered

positive, and are added to 0 degrees to get True North as shown in FIGURE 8-33.

For example, the declination for Logan, Utah is 13.5° East. True North is 360° –

13.5°, or 346.5° as read on a compass.

Orientation is most easily done with two people, one to aim and adjust the

sensor, while the other observes the wind direction displayed by the datalogger.

1. Establish a reference point on the horizon for True North.

2. Sighting down the instrument center line, aim the nose cone, or counterweight

at True North. Display the input location for wind direction using the *6

Mode of the datalogger, or, the Monitor Mode of LoggerNet with an on-line

PC.

3. Loosen the band clamps or set screws that secure the base of the sensor to the

mast or crossarm. While holding the vane position, slowly rotate the sensor

base until the datalogger indicates 0 degrees. Tighten the band clamps or set

screws loosened previously.

4. Engage the orientation ring indexing pin in the notch at the instrument base

(05103, 05106, and 05305 sensors only), and tighten the band clamp on the

orientation ring.

* Other methods employ observations using the North Star or the sun, and

are discussed in the Quality Assurance Handbook for Air Pollution

Measurement Systems, Volume IV - Meteorological Measurements

4

.

47

Advertising