3
IEEE-488 Interface Commands
3.1 IEEE Mandated and Optional Common Commands
The required common commands are IEEE-488.2 mandated commands that are defined in the IEEE-
488.2 standard and must be implemented by all SCPI compatible instruments. These commands are
identified by the asterisk (*) at the beginning of the command keyword. These commands are used to
control instrument status registers, status reporting, synchronization, and other common functions.
Commands declared mandatory by IEEE 488.2.
*CLS Clear Status Command
*ESE Standard Event Status Enable Command
*ESE? Standard Event Status Enable Query
*ESR? Standard Event Status Register Query
*IDN? Identification Query
*OPC Operation Complete Command
*OPC? Operation Complete Query
*RST Reset Command
*SRE Service Request Enable Command
*SRE? Service Request Enable Query
*STB? Read Status Byte Query
*TST? Self-Test Query
*WAI Wait-to-Continue Command
Optional common commands described by IEEE 488.2
*OPT? Option Identification Query
3.1.1
*CLS
The Clear Status (CLS) command clears the status byte by emptying the error queue and clearing all
the event registers including the Data Questionable Event Register, the Standard Event Status
Register, the Standard Operation Status Register and any other registers that are summarized in the
status byte.