21 :system subsystem, Syst, Ubsystem – BNC 835 Programmer Manual User Manual

Page 61: System subsystem

Advertising
background image


Berkeley Nucleonics Corporation 2955 Kerner Blvd., San Rafael, CA 94901

Phone: 415-453-9955, Fax: 415-453-9956, Email: [email protected], Web: www.berkeleynucleonics.com

5.21

:SYSTem Subsystem

Command

Parameters

Unit

(default)

Remark

:SYSTem:ERRor[:NEXT]?

:SYSTem:ERRor:ALL?

:SYSTem:PRESet

:SYSTem:VERSion?

:SYSTem:LOCK

:SYSTem:LOCK:RELease

:ERRor?

:SYSTem:ERRor[:NEXT]?

Return Parameters: Integer error number

Query command is a request for the next entry in the instrument’s error queue. Error messages in the

queue contain an integer in the range [–32768, 32768] denoting an error code and associated

descriptive text.

:ERRor:ALL?

:SYSTem:ERRor:ALL?

Return Parameters: List of integer error number

Query command is a request for the all entries in the instrument’s error queue. Error messages in the

queue contain an integer in the range [–32768, 32768] denoting an error code and associated

descriptive text. This query clears the instrument’s error queue.

:PRESet

:SYSTem:PRESet

Resets most signal generator functions to factory- defined conditions. This command is similar to the

*RST command.

:VERSion?

:SYSTem:VERSion?

Returns the SCPI version number that the instrument software complies with [1999.0]

:LOCK

:SYSTem:LOCK

Locks (disables) front panel control.

:LOCK:RELease

:SYSTem:LOCK:RELease

Advertising