BNC 835 Programmer Manual User Manual

Page 58

Advertising
background image


Berkeley Nucleonics Corporation 2955 Kerner Blvd., San Rafael, CA 94901

Phone: 415-453-9955, Fax: 415-453-9956, Email: [email protected], Web: www.berkeleynucleonics.com

responds (plays) during the high state of the trigger signal. When the BENCHTOP MICROWAVE

SOURCE receives multiple trigger occurrences when only one is required, the signal generator uses

the first trigger and ignores the rest. The Commands NP and PN select both edges with negative first

or positive first.

*RST POS

:TRIGger[:SEQuence]:ECOunt

TRIGGer[:SEQuence]:ECOunt <value>

TRIGGer[:SEQuence]:ECOunt?

This command sets a modulo on consecutive trigger events. Setting the value to N means that only

every Nth trigger event will be considered. Setting it to one means will use every trigger event that

does not occur during a running sweep.

*RST 1

Range 1 to 255

:TRIGger:OUTPut:POLarity

TRIGGer:OUTPut:POLarity NORMal|INVerted

TRIGGer:OUTPut:POLarity?

This command sets the trigger output signal polarity. Note that the trigger output must be enabled by

sending the [:SOURce]:LFOutput:SOURce TRIGger command.

*RST NORM

:TRIGger:OUTPut:MODE

TRIGGer:OUTPut:MODE NORMal|GATE|POINt

This command sets the trigger output signal mode.

NORMal

The trigger output signal is pulsed once whenever playing a waveform

sequence is triggered.

GATE

The trigger output signal is set when playing a waveform sequence is

triggered, and reset when playing stops.

POINT

The trigger output signal is pulsed for each point of the sweep (list) playing.

Note that the trigger output must be enabled by sending the [:SOURce]:LFOutput:SOURce TRIGger

command.

*RST NORM

Advertising