BNC 835 Programmer Manual User Manual
Page 58
Berkeley Nucleonics Corporation 2955 Kerner Blvd., San Rafael, CA 94901
Phone: 415-453-9955, Fax: 415-453-9956, Email: [email protected], Web: www.berkeleynucleonics.com
responds (plays) during the high state of the trigger signal. When the BENCHTOP MICROWAVE
SOURCE receives multiple trigger occurrences when only one is required, the signal generator uses
the first trigger and ignores the rest. The Commands NP and PN select both edges with negative first
or positive first.
*RST POS
:TRIGger[:SEQuence]:ECOunt
TRIGGer[:SEQuence]:ECOunt <value>
TRIGGer[:SEQuence]:ECOunt?
This command sets a modulo on consecutive trigger events. Setting the value to N means that only
every Nth trigger event will be considered. Setting it to one means will use every trigger event that
does not occur during a running sweep.
*RST 1
Range 1 to 255
:TRIGger:OUTPut:POLarity
TRIGGer:OUTPut:POLarity NORMal|INVerted
TRIGGer:OUTPut:POLarity?
This command sets the trigger output signal polarity. Note that the trigger output must be enabled by
sending the [:SOURce]:LFOutput:SOURce TRIGger command.
*RST NORM
:TRIGger:OUTPut:MODE
TRIGGer:OUTPut:MODE NORMal|GATE|POINt
This command sets the trigger output signal mode.
NORMal
The trigger output signal is pulsed once whenever playing a waveform
sequence is triggered.
GATE
The trigger output signal is set when playing a waveform sequence is
triggered, and reset when playing stops.
POINT
The trigger output signal is pulsed for each point of the sweep (list) playing.
Note that the trigger output must be enabled by sending the [:SOURce]:LFOutput:SOURce TRIGger
command.
*RST NORM