5.7

[SOURce]:POWer Subsystem

Command

Parameters

Unit

(default)

Remark

[SOURce]:POWer[:LEVel][:IMMediate][:AMPLitude]

<value>

dBm

[SOURce]:POWer:MODE

FIXed|LIST|SWEep

FIXed

[SOURce]:POWer:STARt

<value>

dBm

[SOURce]:POWer:STOP

<value>

dBm

[SOURce]:POWer:STEP?

dB

[SOURce]:POWer:ALC

ON|OFF|1|0

ON

[SOURce]:POWer:ALC:LOWN

ON|OFF|1|0

OFF

[SOURce]:POWer:ALC:HOLD

ON|OFF|1|0

OFF

[SOURce]:POWer:ATTenuation

<value>

dB

[SOURce]:POWer:ATTenuation:AUTO

ON|OFF|1|0

ON

[SOURce]:POWer:ATTenuation:LIST?

[:LEVel][:IMMediate][:AMPLitude]

[SOURce]:POWer[:LEVel][:IMMediate][:AMPLitude] <value><unit>

[SOURce]:POWer[:LEVel][:IMMediate][:AMPLitude]?

This command sets the RF output power.

:MODE

[SOURce]:POWer:MODE CW|LIST|SWEep

[SOURce]:POWer:MODE?

This command sets the signal generator power mode to fixed or swept.

FIX

This choice stops a power sweep, allowing the signal generator to operate at

a fixed power level. Refer to

[:LEVel][:IMMediate][:AMPLitude]

command

for setting the output power level.

LIST or SWEep

This choice selects the swept power mode. If sweep triggering is set to

immediate along with continuous sweep mode, executing the command starts

the LIST or STEP power sweep.

:STARt

[SOURce]:POWer:STARt <value><unit>

[SOURce]:POWer:STARt?

This command sets the first amplitude point in a step sweep.

Range

:STOP