BNC 835 Programmer Manual User Manual
Page 44
![background image](/manuals/545972/44/background.png)
Berkeley Nucleonics Corporation 2955 Kerner Blvd., San Rafael, CA 94901
Phone: 415-453-9955, Fax: 415-453-9956, Email: [email protected], Web: www.berkeleynucleonics.com
[SOURce]:SWEep:DWELl?
This command sets the dwell time for the current step sweep points. Dwell time is used when
IMMediate is the trigger source. The dwell time is the amount of time the sweep is guaranteed to
pause after setting the frequency and/or power for the current point.
The setting enabled by this command is not affected by signal generator power- on, preset, or *RST.
The variable <value> is expressed in units of seconds (default: 1000 S).
:SWEep:DELay
[SOURce]:SWEep:DELay <value>
[SOURce]:SWEep:DELay?
This command sets the off time for the current step sweep points. Off time is used when IMMediate is
the trigger source. The off time is the amount of time the sweep is guaranteed to pause after setting
the frequency and/or power for the current point.
The setting enabled by this command is not affected by signal generator power- on, preset, or *RST.
The variable <value> is expressed in units of seconds (default: 300 S).
:SWEep:PROGgress
[SOURce]:SWEep:PROGgress?
This query returns the progress of an active sweep, 0.0…1.0.
:SWEep:SPACing
[SOURce]:SWEep:SPACing LINear | LOGarithmic
[SOURce]:SWEep:SPACing?
This command enables the signal generator linear or logarithmic sweep modes. These commands
require the signal generator to be in step mode.
The instrument uses the specified start frequency, stop frequency, and number of points for both linear
and log sweeps.
*RST LIN
:STARt
[SOURce]:POWer:STARt <value><unit>
[SOURce]:POWer:STARt?
This command sets the first amplitude point in a step sweep.
*RST ?
Range Refer to
[:LEVel][:IMMediate][:AMPLitude]
command for the output power ranges.
:STOP
[SOURce]:POWer:STOP <value><unit>
[SOURce]:POWer:STOP?
This command sets the last amplitude point in a step sweep.
*RST ?