Dispensing liquids
epMotion

®

5070

English (EN)

32

6.2

Dispensing modes

The epMotion can be used to dispense liquids in the

Pipette

and

Multidispense

modes.

6.2.1

Dispensing mode:

Pipette

In the

Pipette

mode, a defined volume of liquid is aspirated and fully dispensed.

Dispensing errors are smaller than in the

Multidispense

(see Dispensing tool errors on p. 87) mode.

The epMotion executes a blow-out in the

Pipette

mode. Residual liquid from the pipette tip will be

dispensed in the destination vessel. The blow-out is part of the dispensing volume. The blow-out can be set
in the software (refer to epMotion software operating manual).

In the

Pipette

mode, the smallest dispensing volume for water is 1 μL.

6.2.2

Dispensing mode:

Multidispense

In the

Multidispense

mode, liquid is aspirated and dispensed in the destination vessels in defined partial

quantities.

Dispensing errors are larger than in the

Pipette

(see Dispensing tool errors on p. 87) mode.

The epMotion carries out a remaining stroke in the

Multidispense

mode. Residual liquid from the pipette tip

will be dispensed in the waste vessel or source vessel. The residual liquid is not part of the dispensing
volume.

In the

Multidispense

mode, the smallest dispensing volume for water is 3 μL.

6.2.2.1 Aspiration volume at

Multidispense

To create the same physical conditions for all dispensing steps, more liquid must be aspirated in the

Multidispense

mode than the sum of the dispensing volumes. This is the result of the reverse stroke and the

remaining stroke.

Reverse stroke

• The reverse stroke occurs after liquid aspiration.

Part of the aspirated volume is immediately returned to the source vessel.
The reverse stroke is the same size for all liquids.

Remaining stroke

• After the reverse stroke, there is more liquid in the pipette tip than the amount required for the

dispensing steps. This additional aspiration volume will be dispensed as a remaining stroke after
dispensing.
The additional aspiration volume will be dispensed to the source vessel if the tips are not replaced
before each liquid aspiration.
The additional aspiration volume will be dispensed to the waste container if the tips are not replaced
before each liquid aspiration.