4 thermoadapter frosty, Thermoadapter frosty – Eppendorf epMotion 5070 User Manual

Page 56

Advertising
background image

Labware accessories
epMotion

®

5070

English (EN)

56

8.2.4

Thermoadapter Frosty

Abb. 8-13: Thermoadapter Frosty

Fig. 8-13: Thermoadapter Frosty

The Thermoadapter Frosty cools samples.

The Thermoadapter Frosty is a modified height adapter which includes a PCR-Cooler. A PCR plate is placed
on the Thermoadapter Frosty. The PCR-Cooler cools the samples in this plate. The Thermoadapter Frosty
can only be used with skirted PCR plates.

Preparing and equipping the Thermoadapter Frosty

1. Cool the PCR-Cooler by placing the PCR-Cooler in the lab refrigerator with the upper side down.

2. Insert the PCR-Cooler in the Thermoadapter Frosty.

3. Place a PCR plate on the Thermoadapter Frosty.

The PCR-Cooler changes color if a temperature of 7 °C is exceeded. The color changes from
purple to pink or from dark blue to light blue. The color value in the recesses of the
PCR-Cooler is a decisive factor for cooling samples.

Advertising