81

11.2.2

Dispensing errors

11.2.3

Software errors

The software operating manual contains a description of software error messages.

Symptom/message Cause

Remedy

The dispensing tool
dispenses
incorrectly.

• The sealing rings on the dispensing

tool are broken.

 Replace the sealing rings on the

dispensing tool.

Dispensing volume
too high.

During retraction, the opening of the
pipette tip is in the liquid in the source
vessel. There are no air bubbles on the
lower end of the pipette tip when liquids
are being transported.

• The actual liquid volume in the source

vessel is higher than the entered
volume.

 Use the optical sensor to determine

the volume.

 Enter the correct volume.
 Avoid plates with very different well

filling heights.

• Liquid refilled after the application

was started.

 Do not refill liquid after the

application has started.

• Incorrect labware used.

 Use the labware that was selected in

the application software.

Dispensing volume
too low.

The pipette tip draws in air. There are air
bubbles in the pipette tip.

• The actual liquid volume in the source

vessel is lower than the entered
volume.

 Use the optical sensor to determine

the volume.

 Enter the correct volume.
 Avoid plates with very different well

filling heights.

• Incorrect labware used.

 Use the labware that was selected in

the application software.

Residual liquid in
pipette tip.

• Incorrect volume specification in the

destination vessel.

• After dispensing, the liquid will be

withdrawn from the destination vessel
to the pipette tip.

 Use the optical sensor to determine

the volume.

 Enter the correct volume.
 Avoid plates with very different well

filling heights.

• Incorrect liquid class selected.
• Inappropriate dispensing parameter.

 Select the liquid class that matches

the liquid.

 Optimizing dispensing parameters

(refer to epMotion software operating
manual).

• Liquid wets the pipette tip.
• Liquid tends to foam.

 For several dispensing steps,

exchange the pipette tip more often.

 Optimizing dispensing parameters

(refer to epMotion software operating
manual).