3 autoclaving, 2 decontamination before shipment, 3 servicing the device – Eppendorf epMotion 5070 User Manual

Page 74: Autoclaving, Decontamination before shipment, Servicing the device

Advertising
background image

Maintenance
epMotion

®

5070

English (EN)

74

10.1.3

Autoclaving

Prerequisites

• The temperature in the autoclave is 121 °C.
• 100 kPa overpressure is present in the autoclave.
• Chemical decontamination agents are removed completely.

1. Autoclave tools and labware for 20 min.

2. Rinse the tools and labware with deionized water.

3. Let the tools and labware dry.

10.2

Decontamination before shipment

10.3

Servicing the device

 Have every dispensing tool serviced after 100,000 full strokes or 200,000 strokes.

 If you radiate your device regularly with UV light, you habe to service this device every year.

NOTICE! Damage on the tool due to autoclaving.
The tool will be damaged if the temperature in the autoclave is too high.

 Make sure the temperature in the autoclave is 121 °C – 126 °C.
 Make sure the tools do not touch the wall of the autoclave.

WARNING! Risk to health from contaminated device

1. Follow the instructions in the decontamination certificate. It is available as a PDF file on

our webpage (www.eppendorf.com/decontamination).

2. Decontaminate all the parts you would like to dispatch.
3. Include the fully completed decontamination certificate in the package.

The Eppendorf AG service team is available to service and certify your device.

The service provisions include:

• Installation qualification (IQ) and operational qualification (OQ)
• Preventive maintenance
• Calibration and validation

Information on the services offered can be found on our webpage: www.eppendorf.com/
epservices.