8 dispensing tool errors, 1 pipetting, Dispensing tool errors 13.8.1 – Eppendorf epMotion 5070 User Manual

Page 87: Pipetting

Advertising
background image

Technical data

epMotion

®

5070

English (EN)

87

13.8

Dispensing tool errors

13.8.1

Pipetting

Dispensing tool

Volume range
Pipetting

Testing
volume

Error limits

Error

Systematic error

Random error

± %

± μL

± %

± μL

TS 50

1.0 μL – 50 μL

1 μL
5 μL
25 μL
50 μL

±15.0
±5.0
±1.5
±1.0

±0.15
±0.25
±0.375
±0.5

±5.0
±3.0
±0.6
±0.4

±0.05
±0.15
±0.15
±0.2

TS 300

20 μL – 300 μL

20 μL
30 μL
150 μL
300 μL

±4.0
±3.0
±1.0
±0.6

±0.8
±0.9
±1.5
±1.8

±2.5
±1.5
±0.4
±0.3

±0.5
±0.45
±0.6
±0.9

TS 1000

40 μL – 1 000 μL

40 μL
100 μL
500 μL
1 000 μL

±5.0
±2.0
±1.0
±0.7

±2.0
±2.0
±5.0
±7.0

±1.5
±1.0
±0.2
±0.15

±0.6
±1.0
±1.0
±1.5

TM 50-8

1.0 μL – 50 μL

1 μL
5 μL
25 μL
50 μL

±25.0
±5.0
±2.0
±1.2

±0.25
±0.25
±0.5
±0.6

±10.0
±5.0
±1.2
±0.6

±0.1
±0.25
±0.3
±0.3

TM 300-8

20 μL – 300 μL

20 μL
30 μL
150 μL
300 μL

±10.0
±10.0
±2.5
±1.5

±2.0
±3.0
±3.75
±4.5

±4.0
±3.5
±0.8
±0.5

±0.8
±1.05
±1.2
±1.5

TM 1000-8

40 μL – 1 000 μL

40 μL
100 μL
500 μL
1 000 μL

±6.0
±3.0
±1.5
±0.8

±2.4
±3.0
±7.5
±8.0

±2.5
±1.5
±0.3
±0.15

±1.0
±1.5
±1.5
±1.5

Advertising