1 carrying out wipe decontamination, Carrying out wipe decontamination – Eppendorf epMotion 5070 User Manual

Page 72

Advertising
background image

Maintenance
epMotion

®

5070

English (EN)

72

10.1.1

Carrying out wipe decontamination

Prerequisites

• Decontamination agent
• Deionized water
• Lint-free cloths
• Cotton swabs

Cleaning the device

1. Switch off the device and disconnect it from the mains/power line.

2. Moisten the lint-free cloth and cotton swab with decontamination agent.

3. Clean the surface with a lint-free cloth. Clean hard to access surfaces with cotton swabs.

The surfaces are dampened with a decontamination agent.

4. Allow the decontamination agent to take effect.

5. Wipe the decontamination agent with deionized water.

6. Allow the surfaces to dry.

Cleaning the dispensing tools

7. Pull of the ejector from the single-channel dispensing tool.

Spray decontamination
outside the device

70% (v/v) ethanol

Waste box
LiquidWasteTub
Racks
Thermoblocks
Thermoadapter
Height adapter
TipHolder

UV radiation via Cleanbench

UV light of the Cleanbench
Irradiation period 15 min
If you are using UV irradiation
regularly, the epMotion must
be serviced annually by
Eppendorf Service.

Worktable without labware and tools

Autoclaving

121 °C
100 kPa overpressure
20 min

Waste box
LiquidWasteTub
Dispensing Tools
Racks
Thermoblocks
Thermoadapter
Height adapter
TipHolder

Decontamination method

Decontamination agent

For use on

Advertising