10

Maintenance

10.1

Decontamination and cleaning

The device is cleaned as described in the Decontamination section.

Tab. 10-1: Decontamination methods and decontamination agents

WARNING! Danger due to flammable and infectious liquids in the waste container.
Pipette tips in the waste container may contain flammable or infectious liquids.

 Wear personal protective equipment.
 Handle the pipette tips and sample materials from the waste container in accordance with

the material safety data sheets, safety regulations and laboratory guidelines.

WARNING! Danger due to contamination of the LiquidWasteTub.
The outer surface of the LiquidWasteTub may become contaminated by pipette tips in the
waste container.

 Wear personal protective equipment.
 Decontaminate the outer surface of the LiquidWasteTub.

NOTICE! Material damage from corroding cleaning agents.
Cleaning and decontamination agents may contain corroding substances. Metal surfaces on
the epMotion become damaged by corrosion.

 Use the cleaning agents and disinfectants specified in the operating manual.

Decontamination method

Decontamination agent

For use on

Wipe decontamination

70% (v/v) ethanol

Worktable
Waste container
LiquidWasteTub
Housing
Work surface adapter
Reflectors and sensor cover on the
Cleanbench
EasyCon Control-Panel
Labware
Tools

3%–4% sodium hypochlorite
solution

Waste container
LiquidWasteTub
Dispensing Tools
Racks
Thermoblocks
Thermoadapter
Height adapter
TipHolder