4 fuses, 5 carrying out firmware updates, 6 checking the dispensing volume – Eppendorf epMotion 5070 User Manual

Page 75: Fuses, Carrying out firmware updates, Checking the dispensing volume

Advertising
background image

Maintenance

epMotion

®

5070

English (EN)

75

10.4

Fuses

Fuses may only be replaced by authorized service technicians. Users must not replace the fuses.

10.5

Carrying out firmware updates

Prerequisites

• The firmware is located on a USB storage medium.
• The device is switched off and disconnected from the mains/power line.

1. Connect the USB storage medium to a USB port on the rear of the epMotion.

2. Switch on the epMotion.

The epMotion automatically installs the firmware.

3. Wait until the epMotion has emitted 3 short signal tones.

4. Exit epBlue (refer to epMotion software operating manual).

The EasyCon switches off. The display becomes dark.

5. Switch off the epMotion.

6. Remove the USB storage medium.

10.6

Checking the dispensing volume

When you use liquids with a density greatly different from that of water, the dispensing volume may be
incorrect. Physical properties, such as the viscosity, vapor pressure and surface tension of the liquid, affect
the dispensing volume.

Correct the dispensing volume for these liquids. Gravimetrically check the dispensing of the epMotion and
correct the dispensing if necessary.

Perform the following check.

1. Call up the application.

2. Adjust the application to your labware and dispensing tool.

3. Fill the source vessel with distilled water.

4. Weigh empty labware.

5. Run the application.

The labware will be filled with water.

6. Weigh filled labware.

7. Calculate the volume of the water. Volume = mass : density.

NOTICE! Malfunctions due to faulty firmware.

 Firmware updates only be conducted upon request by Eppendorf AG.
 Only use firmware you have received from Eppendorf AG.