3.2 Heater/Stirrer Set-Up and Programming

3.2a General Information

Circulator Pump

Pump speed is selected via the Main Menu.

The MINIMUM setting is used for DGGE applications.

MAXIMUM is recommended where temperature varies frequently and there is a need for fast recovery or when

pumping to multiple external units.

Speed

Maximum Pump Outlet Ratings

Maximum Pump Outlet Ratings

Selection

Line Frequency = 60Hz

Line Frequency = 50Hz

MAXIMUM

30 LPM / 5.0 PSI

22 LPM / 3.4 PSI

HIGH

24 LPM / 2.5 PSI

20 LPM / 1.7 PSI

MEDIUM

20 LPM / 1.7 PSI

17.5 LPM / 1.2 PSI

LOW

18 LPM / 1.2 PSI

15.8 LPM / 0.8 PSI

MINIMUM

15.7 LPM / 0.8 PSI

15 LPM / 0.7 PSI

The data in the table above are based on the following criteria:
a. Maximum pump outlet flow rate is measured in liters per minute (LPM) with no restriction on the pump outlet.
b. Maximum pump outlet pressure is measured in pounds per square inch (PSI) at no flow.
c. Water was used as the circulation fluid. Water has a viscosity of one centistoke. High viscosity or low-density fluids will change

these figures.

Filling the Reservoir
A liquid level that fully covers the heater coil, pump, over-temperature sensor, and at least one inch

(25mm) of the temperature sensor must be maintained. For heating applications, the TTGE tank

must

have at least 23 liters of buffer. When running the system with two dual cassettes, buffer volume should

be 24 liters.

minimum fill level

Temperature Sensor

Model Type

Temperature

Range

Reservoir

Capacity

Amps @

120V, 60Hz

Amps @

240V, 50Hz

Heat Only

Immersion Circulator

Ambient +5°C to

200°C*

N/A

11A

9.7A