.2 Heater/Stirrer Set-Up and Programming – continued

3.2r Setting Preferences

The Preferences sub-menu allows you program global preferences regarding instrument operation.

Readout — This is the number of decimal places to which temperatures will be displayed (0, 1, or 2).

Units — This is the unit in which temperatures will be displayed (°C or °F)

Sound — This is the volume level for the unit’s audible signal. When it is selected, the volume of the audible signal

changes as the Select/Set Knob is rotated. You must press the Select/Set Knob to accept the displayed volume

value; if you press the ESC Button or allow the display to timeout without pressing the Select/Set Knob, the sound

level will remain where it was previously set.

Language — This is the language used for displays. When this is selected, a sub-menu appears with the available

languages.

Program — This menu item appears only on the Programmable Controller. It is used to select whether programs are

run using Time or Temperature as the priority.
When Time is selected, the program begins running when the bath temperature reaches the programmed set point

for step one and continues until the total programmed period of time has elapsed. The set point target for any given

step (except step one) may or may not be reached before the program advances to the next step.

When Temperature is selected as the priority, the program begins running when the bath temperature reaches

the programmed set point for step one. Each subsequent step runs until the programmed set point for that step is

reached, regardless of how much time has elapsed.

Advertising